Model 979 Series Helium Mass Spectrometer Leak Detector
DRAFT 9/15/03
2-13
GROSS LEAK
GROSS LEAK: INSTALLED
Gross leak installed–standard on dual
pump configurations.
GROSS LEAK: NOT INSTALLED
Gross leak not installed.
ION SOURCE FILAMENT
FIL 1 OFF
Filament 1 is selected and not lit.
FIL 1 ON
Filament 1 is selected and lit.
FIL 2 OFF
Filament 2 is selected and not lit.
FIL 2 ON
Filament 2 is selected and lit.
LEAK RATE RANGE
AUTO RANGING
System is set up to auto range through
leak rate scale.
MANUAL RANGING
System is set up for manual range
control.
RANGE STOP
RANGE STOP -[EXP]
System is set up so that the most
sensitive leak rate range displayed is
10
-
[EXP]
.
REJECT SET POINT
REJECT# 1,2,3,4 ACTIVE
Reject set points displayed are active;
this line is blank when all set points are
disabled or inactive.
ROUGH PUMP (Dual pump
configuration only)
ROUGH PUMP: ROUGH ONLY
Dedicated rough pump is connected to
the test port only during the roughing
cycle.
ROUGH PUMP: SPLIT FLOW
Dedicated rough pump is connected to
the test port during rough and test
cycles, providing additional pumping
capacity during test.
SYSTEM SENSITIVITY
HIGH SENS
Leak detector is configured for high
sensitivity testing. This line is blank
when the leak detector is configured for
standard sensitivity testing.
Table 2-4
System Information Screen Conditions (Continued)
CONDITION
DISPLAY
DESCRIPTION
Содержание 979 Series
Страница 12: ...Model 979 Series Helium Mass Spectrometer Leak Detector This page intentionally left blank D R A F T 9 1 5 0 3 ...
Страница 112: ...Model 979 Series Helium Mass Spectrometer Leak Detector This page intentionally left blank D R A F T 9 1 5 0 3 ...
Страница 132: ...Model 979 Series Helium Mass Spectrometer Leak Detector This page intentionally left blank D R A F T 9 1 5 0 3 ...
Страница 138: ...Model 979 Series Helium Mass Spectrometer Leak Detector This page intentionally left blank D R A F T 9 1 5 0 3 ...
Страница 139: ......