Model 979 Series Helium Mass Spectrometer Leak Detector
DRAFT 9/15/03
2-23
Table 2-5
979 Standard Sensitivity Leak Detector
Test
Fine
Gross (2 Pump)
Range
–9
–8
–7
–6
–5
–4
–3
–2
–1
Valve Path
MS*
MS
MS/CF
MS/CF
CF**
CF/GL GL***
GL
GL
Transfer Pressure
100 mTorr 100 mTorr
100 mTorr /
5 Torr
100 mTorr /
5 Torr
5 Torr
5 Torr
atm
atm
atm
*MS – MIDSTAGE
**CF – CONTRA-FLOW
***GL – GROSS LEAK
Table 2-6
979 High Sensitivity Leak Detector
Test
Fine
Gross (2 Pump)
Range
–10
–9
–8
–7
–6
–5
–4
–3
–2
Valve Path
MS
1
MS
MS/CF
MS/CF
CF
2
CF/GL
CF/GL
GL
3
GL
Transfer Pressure
1 Pump
4
100 mTorr 100 mTorr
100 mTorr
/ 5 Torr
100 mTorr
/ 5 Torr
5 Torr
5 Torr
Transfer Pressure
2 Pump
4
100 mTorr 100 mTorr
100 mTorr
/ 2 Torr
100 mTorr
/ 2 Torr
2 Torr 2 Torr / atm
atm
atm
atm
1
MS – MIDSTAGE
2
CF – CONTRA-FLOW
3
GL – GROSS LEAK
4
Contra-flow tranfers settings are as follows:
❑
Single pump systems (no gross test): 5 Torr max.
❑
Two pumps (has gross test): 2 Torr max.
Содержание 979 Series
Страница 12: ...Model 979 Series Helium Mass Spectrometer Leak Detector This page intentionally left blank D R A F T 9 1 5 0 3 ...
Страница 112: ...Model 979 Series Helium Mass Spectrometer Leak Detector This page intentionally left blank D R A F T 9 1 5 0 3 ...
Страница 132: ...Model 979 Series Helium Mass Spectrometer Leak Detector This page intentionally left blank D R A F T 9 1 5 0 3 ...
Страница 138: ...Model 979 Series Helium Mass Spectrometer Leak Detector This page intentionally left blank D R A F T 9 1 5 0 3 ...
Страница 139: ......