4-5
Model 979 Series Helium Mass Spectrometer Leak Detector
DRAFT 9/15/03
CAUTION
Use powder-free butyl or polycarbonate gloves to prevent skin
oils from getting on vacuum surfaces.
4.3.1 Removing the Spectrometer Tube Assembly
WARNING
Disconnect power from the 979 before performing any
maintenance procedure that requires physically disconnecting
any part of the system.
The spectrometer tube operates at a very high vacuum produced by the high vacuum pump.
Service of the spectrometer tube requires that this vacuum be vented to the atmosphere.
Perform this procedure every twelve months, or sooner if required. Access to the spectrom-
eter tube is gained by removing the 2 Phillips screws from the front panel and dropping the
top of the front panel down.
NOTE
Rebuilt spectrometer tubes are available from Vacuum
Technologies on an exchange basis. Contact the Vacuum
Technologies Service Center (1-800-8VARIAN) for details.
Table 4-3
Tools and Parts Required for Spectrometer Tube Overhaul
Tools
Screwdrivers Slotted and Phillips
Gloves
Butyl or polycarbonate, powder-free
Parts
Part Number
Description
82850302
Ion Source
R1266301
Button TC
670029096
Vacuum Technologies Spectrometer Tube Cleaning Kit
O-rings (Parker 2-025 V747-75 Black) – (supplied with Ion Source, Button TC and
Preamplifier)
Содержание 979 Series
Страница 12: ...Model 979 Series Helium Mass Spectrometer Leak Detector This page intentionally left blank D R A F T 9 1 5 0 3 ...
Страница 112: ...Model 979 Series Helium Mass Spectrometer Leak Detector This page intentionally left blank D R A F T 9 1 5 0 3 ...
Страница 132: ...Model 979 Series Helium Mass Spectrometer Leak Detector This page intentionally left blank D R A F T 9 1 5 0 3 ...
Страница 138: ...Model 979 Series Helium Mass Spectrometer Leak Detector This page intentionally left blank D R A F T 9 1 5 0 3 ...
Страница 139: ......