Models 947, 948 and 960 Leak Detectors Operations and Service Manual
3-34
7. Press and hold the
VENT
button for approximately 1 second until the unit switches to
the VENT mode.
❑
The TEST LED turns off.
❑
The VENT LED illuminates.
NOTE
If the VENT button is pressed for less than one second, the leak
detector moves to HOLD.
8. Remove the tested part from the test port.
The Leak test is now complete.
3.15 Standby
When continued use of 947/948/960 is not anticipated, STANDBY
mode, a partial
shutdown, can be used to save energy. To put the 947/948/960 into STANDBY
mode:
1. Press the
VENT
button and hold it until the Vent LED comes on.
2. Place the Rough pump power switch in the
OFF
position.
3. Turn the Rough pump switch breaker to the
ON
position to return the Leak Detector to
the READY FOR OPERATION mode.
NOTE
When the unit is returned to Ready for Operation from Standby,
the warmup periods do not apply.
3.16 Shutdown
Complete shutdown for a move or temporary storage is accomplished by the following
shutdown procedure:
1. Place the test port plug securely in the test port.
2. Turn the ROUGH PUMP switch
OFF
.
3. Turn DETECTOR switch
OFF
.
4. Turn the Power switch/circuit breaker
OFF
on back of cabinet.
5. For the 948/960 only, empty the cold trap.
See “Liquid Nitrogen Cold Trap (Models 948/960 only)” on page 4-11 for the cleaning
procedure.
Содержание 960
Страница 12: ...Models 947 948 and 960 Leak Detectors Operations and Service Manual This page intentionally left blank...
Страница 32: ...Models 947 948 and 960 Leak Detectors Operations and Service Manual xxxii...
Страница 74: ...Models 947 948 and 960 Leak Detectors Operations and Service Manual This page intentionally left blank...
Страница 148: ...Models 947 948 and 960 Leak Detectors Operations and Service Manual This page intentionally left blank...
Страница 156: ...Models 947 948 and 960 Leak Detectors Operations and Service Manual 5 8...
Страница 194: ...Models 947 948 and 960 Leak Detectors Operations and Service Manual This page intentionally left blank...
Страница 202: ...Models 947 948 and 960 Leak Detectors Operations and Service Manual This page intentionally left blank...
Страница 216: ...Models 947 948 and 960 Leak Detectors Operations and Service Manual This page intentionally left blank...
Страница 229: ......