Models 947, 948 and 960 Leak Detectors Operations and Service Manual
A-8
[SHIFT]
SET
STD LK
SHOW
Sets the leak rate value to which the leak detector calibrates. Enter
X.XE-X (the desired value), followed by the SHIFT key, then the
function key.
REJ3
SET
REJ 1
Sets the leak rate value for Reject Set Point #1. Enter X.XE-X
(the desired value), then the function key.
[SHIFT]
REJ3
SET
REJ 1
Sets the leak rate value for Reject Set Point #3. Enter X.XE-X
(the desired value), followed by the SHIFT key, then the
function key.
REJ 4
SET
REJ 2
Sets the leak rate value for Reject Set Point #2. Enter X.XE-X
(the desired value), followed by the SHIFT key, then the
function key.
[SHIFT]
REJ 4
SET
REJ 2
Sets the leak rate value for Reject Set Point #4. Enter X.XE-X
(the desired value), followed by the SHIFT key, then the
function key.
Содержание 960
Страница 12: ...Models 947 948 and 960 Leak Detectors Operations and Service Manual This page intentionally left blank...
Страница 32: ...Models 947 948 and 960 Leak Detectors Operations and Service Manual xxxii...
Страница 74: ...Models 947 948 and 960 Leak Detectors Operations and Service Manual This page intentionally left blank...
Страница 148: ...Models 947 948 and 960 Leak Detectors Operations and Service Manual This page intentionally left blank...
Страница 156: ...Models 947 948 and 960 Leak Detectors Operations and Service Manual 5 8...
Страница 194: ...Models 947 948 and 960 Leak Detectors Operations and Service Manual This page intentionally left blank...
Страница 202: ...Models 947 948 and 960 Leak Detectors Operations and Service Manual This page intentionally left blank...
Страница 216: ...Models 947 948 and 960 Leak Detectors Operations and Service Manual This page intentionally left blank...
Страница 229: ......