Models 947, 948 and 960 Leak Detectors Operations and Service Manual
4-14
6. Disconnect all four electrical connections to the spectrometer tube
➀
.
NOTE
To remove the locking connector shown in Figure 4-2,
both metal tabs at the left of the connector must be pushed to
unlock the connector, then pull the connector to the right to
remove it.
Figure 4-2
947/948/960 Spectrometer Tube Removal
7. Hold the spectrometer tube and loosen the wing nut holding the KF clamp
➁
.
8. Remove the clamp that holds the spectrometer tube to the Service Block Assembly
(SBA).
9. Rock the tube to slowly break any residual vacuum.
10. Place the spectrometer tube and O-ring on a clean, non-magnetic bench for further
maintenance.
➀
➁
➂
Содержание 960
Страница 12: ...Models 947 948 and 960 Leak Detectors Operations and Service Manual This page intentionally left blank...
Страница 32: ...Models 947 948 and 960 Leak Detectors Operations and Service Manual xxxii...
Страница 74: ...Models 947 948 and 960 Leak Detectors Operations and Service Manual This page intentionally left blank...
Страница 148: ...Models 947 948 and 960 Leak Detectors Operations and Service Manual This page intentionally left blank...
Страница 156: ...Models 947 948 and 960 Leak Detectors Operations and Service Manual 5 8...
Страница 194: ...Models 947 948 and 960 Leak Detectors Operations and Service Manual This page intentionally left blank...
Страница 202: ...Models 947 948 and 960 Leak Detectors Operations and Service Manual This page intentionally left blank...
Страница 216: ...Models 947 948 and 960 Leak Detectors Operations and Service Manual This page intentionally left blank...
Страница 229: ......