Models 947, 948 and 960 Leak Detectors Operations and Service Manual
3-23
REJECT SET POINTS
The REJECT SET POINTS are used to alert the operator that a mea-
sured leak rate has exceeded the set value. There are four set points
available. Each set point must be enabled before they are opera-
tional. Conversely, each must be disabled to stop its operation.
Only one set point can be active at any given time.
AUDIO VOLUME
The AUDIO VOLUME control is used to activate the audio tone
from the turret. The audio tone frequency increases as the size of
the measured leak increases. The audio volume control is located
on the turret.
BAR GRAPH DISPLAY The 50-segment bar graph display on the front control panel repre-
sents the mantissa value of the leak rate, spanning from 0 to 10.
The scale shows above the bar graph. The numerical display,
located to the far right of the bar graph, displays the leak rate expo-
nent value in atm cc/sec.
LINEAR VS LOG
OUTPUT
Configure the analog output voltage to the optional I/O port on the
back of the unit. The options for output are linear or logarithmic at
2 VDC per decade or 3 VDC per decade.
AUTO-SEQUENCER
Provides clear accept/reject signals and allows programming for
roughing, testing, and cycling times. Auto sequencer provides the
ability to set up the 947/948/960 so that it automatically runs
through a complete test cycle (START/ROUGH/TEST/VENT), pass-
ing or failing the test part per the test specification entered by the
user. The auto sequencer set-up controls include test object rough
time, test time, and auto sequencer on/off. A Pass or Fail condition
is determined based upon the reject set point selected by the user
and/or achievement of the pressure transfer set points within the
period of time specified in Rough Time.
The auto sequencer pass or fail leak rate value is selected through
the Reject Set Point. The pass/fail criterion is based on the reject set
point that is enabled (ON). If more than one set point is enabled,
then the pass/fail criterion is based on the most stringent reject set
point value.
Содержание 960
Страница 12: ...Models 947 948 and 960 Leak Detectors Operations and Service Manual This page intentionally left blank...
Страница 32: ...Models 947 948 and 960 Leak Detectors Operations and Service Manual xxxii...
Страница 74: ...Models 947 948 and 960 Leak Detectors Operations and Service Manual This page intentionally left blank...
Страница 148: ...Models 947 948 and 960 Leak Detectors Operations and Service Manual This page intentionally left blank...
Страница 156: ...Models 947 948 and 960 Leak Detectors Operations and Service Manual 5 8...
Страница 194: ...Models 947 948 and 960 Leak Detectors Operations and Service Manual This page intentionally left blank...
Страница 202: ...Models 947 948 and 960 Leak Detectors Operations and Service Manual This page intentionally left blank...
Страница 216: ...Models 947 948 and 960 Leak Detectors Operations and Service Manual This page intentionally left blank...
Страница 229: ......