Varian ELSD Service Manual
Unpacking the detector
Care has been taken to ensure that the instrument should be received in proper condition. The packing and
protection are designed for normal hazards of road, rail or air transit. Any damage to the container or instrument
should be reported immediately to your local distributor, or to Varian. It is recommended that the shipping container
be kept, if possible, for re-shipment or return to a service centre.
Examine the shipping carton for visible signs of exterior damage. Unpack the instrument and examine it for transit
damage. Check that all items on the packing list are included.
Notify your local distributor or Polymer Laboratories, Varian of any damage or missing items.
Packing list
•
Varian ELS Detector
•
Operation Manual
•
Control Software[SMB1]
•
Mains leads (UK, EUR, USA)
•
RS232 data cable
•
Analogue output Cable
•
Gas Tubing (2m)
•
PVC Exhaust hose (2m)
•
Tygon SE200 waste hose (7cm)
•
500ml Solvent waste container
Power Connections
•
Before connecting the power cable, ensure the instrument voltage rating matches your local power supply.
•
Use only a supply with protective grounding.
•
The correct fuses should be installed.
•
For 115V (AC) or 230V (AC) use two 250V H 2A T fuses
THIS UNIT IS DOUBLE - FUSED
.
RISK OF FIRE, REPLACE FUSES AS MARKED!
If the voltage rating and fuses are correct for your power source, connect the power cable
Gas Connection
The instrument should be supplied with clean, dry nitrogen gas at a minimum head pressure of 60psi. A 4mm push-
in connector is provided at the rear of the instrument for a convenient connection to the gas source.
To prevent against unnecessary gas usage, an automatic but controlled gas shut off valve is integrated into the gas
inlet manifold. This will only allow gas to pass into the instrument when the instrument is operating. Should the
instrument default to a standby mode the gas valve will close after 15 minutes.
Содержание 380-LC
Страница 44: ... 6 27401 40 Varian ELSD Service Manual Remove the two screws to take out and replace the gas valve ...
Страница 76: ... 6 27401 72 Varian ELSD Service Manual Figure 1 Terminal program Figure 2 Communications dialog ...
Страница 78: ... 6 27401 74 Varian ELSD Service Manual Figure 5 Selecting the HEX file Figure 6 Downloading the Program ...
Страница 79: ... 6 27401 75 Varian ELSD Service Manual Figure 7 Programming the EPROM ...
Страница 81: ... 6 27401 77 Varian ELSD Service Manual Figure 8 ELSD Device ID Terminal program Figure 9 Communications dialog ...
Страница 82: ... 6 27401 78 Varian ELSD Service Manual Figure 10 Successful Communication with Get Data command ...
Страница 83: ... 6 27401 79 Varian ELSD Service Manual Figure 11 Device ID displayed ...