#6/27401
27
Varian ELSD Service Manual
Replacing The Vapour Sensors
When required:
Faulty Vapour sensor(s)
Tools required:
Access key
Small
screwdriver
Parts required
Vapour Sensor (# PL0890-0510)
The Varian ELSD is fitted with a differential vapour sensor
system. The external sensor is located at the rear of the
instrument; the internal sensor is located on the main control
board
Open the detector’s side panel
using the access key, as
shown.
Locate the external vapour
sensor on the rear of the
instrument.
Содержание 380-LC
Страница 44: ... 6 27401 40 Varian ELSD Service Manual Remove the two screws to take out and replace the gas valve ...
Страница 76: ... 6 27401 72 Varian ELSD Service Manual Figure 1 Terminal program Figure 2 Communications dialog ...
Страница 78: ... 6 27401 74 Varian ELSD Service Manual Figure 5 Selecting the HEX file Figure 6 Downloading the Program ...
Страница 79: ... 6 27401 75 Varian ELSD Service Manual Figure 7 Programming the EPROM ...
Страница 81: ... 6 27401 77 Varian ELSD Service Manual Figure 8 ELSD Device ID Terminal program Figure 9 Communications dialog ...
Страница 82: ... 6 27401 78 Varian ELSD Service Manual Figure 10 Successful Communication with Get Data command ...
Страница 83: ... 6 27401 79 Varian ELSD Service Manual Figure 11 Device ID displayed ...