#6/27401
48
Varian ELSD Service Manual
To replace the cooled evaporator
assembly carefully connect the ends of
evaporator tube into the optical block
and nebuliser chamber, as shown.
Reconnect the electrical and gas
connections
DO NOT TIGHTEN ANY SCREWS AT
THIS POINT AS THIS CAN DAMAGE
THE ASSEMBLY
Having positioned the evaporator
assembly, tighten the nebuliser
chamber screws. (shown in red)
This will lock the ends of the assembly
into position and guarantee the
alignment is correct.
Position the mounting bracket
underneath the evaporator assembly.
Whilst pushing the bracket and
evaporator assembly backwards,
towards the side wall of the ELSD,
tighten the evaporator assembly
screws, (shown in yellow).
This will lock the position of the
assembly to the bracket and prevent
rotation of the assembly.