XRpad2 4343
44
www.vareximaging.com
U
SER
M
ANUAL
Table 14
Recommended Separation Distance between Portable and Mobile RF-Communication
Equipment and the X-ray Detector
Recommended Separation Distance between Portable and Mobile RF-Communication Equipment
and the X-ray Detector
The X-ray detector is intended for use in the electromagnetic environment specified below. The installer,
X-ray system manufacturer, or user of the X-ray detector should assure that it is used in such an
environment.
Rated Maximum Output
Power of the
Transmitter (W)
150 kHz to 80 MHz
80 MHz to 800 MHz
800 MHz to 2.5 GHz
0.01
0.12
0.12
0.23
0.1
0.38
0.38
0.73
1
1.2
1.2
2.3
10
3.8
3.8
7.3
100
12
12
23
For a transmitter rated at a maximum output power not listed above, the separation distance can be
estimated using the equation in the corresponding column, where
P
is the maximum output (power
rating of the transmitter in watt [W]) according to the transmitter manufacture and
d
is the
recommended separation distance in meter (m).
Note:
This guideline may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects, and people.
P
d
2
.
1
P
d
2
.
1
P
d
3
.
2
Version
01
2017-11-03