includes the specification for the XRpad IPU-2.
Table 11
Specification of the XRpad IPU-2
Electrical Specification
AC Input Voltage [16]
100 V … 240 V
AC Frequency [16]
50 Hz/60 Hz
DC Output [8]
13.5 V/5 A, 15 V/1 A
(voltage level is dependent of load)
Trigger In Signal from Hand Switch [6]
5 V … 24 V/10 mA (SELV)
Trigger Out Signal to Generator [5]
Same level as Trigger In Signal
Trigger In Signal [7, 9]
5 V (SELV)
Trigger Out Signal [7, 10]
5 V
DC Output 5PF [7]
5 V/40 mA
Protection Class:
Mechanical Specification
Size
260 mm (L) × 205 mm (W) × 50 mm (H)
Temperature Ranges
Operating
10°C to 35°C
Transportation/Storage
–10°C to 70°C
Relative Humidity
Operating
10% to 90%
Transportation/Storage
0% to 90%
Ingress Protection Rating
IP40 rated (protection against particles > 1 mm)
WARNING
All external signals that are connected to the XRpad IPU-2 (especially
PREP/EXPOSE and Trigger signals) should be from a Separated or
Safety Extra-Low Voltage (SELV) circuit. Ignoring this warning may
result in electric shock, which may result in severe personal injury,
death, or substantial product damage.
WARNING
The detector must not be used in a situation where it controls the X-ray
dose delivered to the patient.
WARNING
The trigger signals from the detector must not be the sole activation
means for X-ray delivery, and must be used in association with other
control inputs: for example, a hand switch, safety interlocks, and a
means of controlling the beam duration.