Inspection and Maintenance
It is important that the X-ray detector is used safely and as intended. Inspect the detector and its
accessories before use. If any problem is found during the inspection, correct the problem, and
take the measures indicated in this section. The X-ray detector and the XRpad IPU-2
Interface and Power Unit do not have any internal exchangeable parts. Only the accessories
listed in
can be changed after consulting with your establishment’s safety representative.
If the problem cannot be corrected, contact your dealer, distributor, device manufacturer, or any
Varex subsidiaries (regional service headquarters) listed on the last page of this
document.
We recommend that records of the inspection be kept close to the detector. You can use copies of
the checklist in this section, or you can make your own copies of the checklist.
11.1
Daily Inspection
Perform the following inspection daily. If there is any problem, immediately ask your
establishment’s safety representative to contact your dealer, distributor, or device manufacturer.
11.1.1
Before Turning On the Power
WARNING
The X-ray detector must be repaired by Varex authorized personnel
only. Ignoring this warning may result in explosion, fire, electric
shock, or unknown hazards, which may result in severe personal
injury, death, or substantial product damage.
Caution
Inspect the X-ray detector before use. In addition, carry out prescribed,
regular inspections per the instructions in this manual.
Inspection
Result
Remedy
Date
/
Date
/
Date
/
Cables
Check all cables, if
applicable, (Power and
communication tethered
cord, DC-cable, Ethernet
cable, Sync cable) to ensure
that they are not damaged
and the insulation is not
damaged.
Good/Bad Good/Bad Good/Bad Contact your dealer,
distributor, or device
manufacturer if there is a
problem.
Check all connector plugs
and locks to ensure they are
not loose.
Good/Bad Good/Bad Good/Bad Fully insert the cables and
lock them.
Version
01
2017-11-03