Performance Checks
3-4
SurgiStat II-20 User’s Guide
5.
Press the Coag pedal on the footswitch. Verify that the Desiccate, Fulgurate,
and Bipolar Activation Indicator illuminates and that the system generates the
Coag activation tone.
6.
While activating the Coag mode, rotate the volume control over the full range
to verify that the sound is audible throughout the range.
Checking Monopolar Mode (with Handswitch)
1.
Connect a handswitching handpiece to the Monopolar Handswitching
receptacle.
2.
Activate, one at a time, the Cut and Coag handswitching controls.
3.
Verify that each control causes the correct indicator to illuminate and tone to
sound.
Performance Checks
After the unit has passed the preliminary functional test, it is ready for
performance testing. A qualified biomedical engineer who is thoroughly familiar
with electrosurgical devices should conduct this testing. The testing should
include checking all modes of operation for proper function and power output.
For a detailed description of performance testing procedures, refer to the
SurgiStat II Electrosurgical Generator Service Manual
.