Standards and IEC Classifications
T
e
chnical S
p
ecifications
SurgiStat II-20 User’s Guide
8
-
5
•
When the alarm condition is resolved, the green single or split-plate
indicator will illuminate.
•
The system measures the return electrode sensing current according to
IEC 60601-1.
Low Frequency (50–60 Hz) Leakage Current
High Frequency (RF) Leakage Current
Standards and IEC Classifications
The SurgiStat II-20 generator meets all pertinent clauses of the IEC 60601-1
second edition and IEC 60601-2-2 third edition.
Class I Equipment (IEC 60601-1)
Accessible conductive parts cannot become live in the event of a basic insulation
failure because of the way in which they are connected to the protective earth
conductor.
Enclosure source
current, ground open
< 300 µA
Source current, patient
leads, all outputs
Normal polarity, intact ground: < 10 µA
Normal polarity, ground open: < 50 µA
Reverse polarity, ground open: < 50 µA
Sink current at high
line, all inputs
< 50 µA
Bipolar RF leakage
current
< 39 mA
rms
Monopolar RF leakage
current
(additional tolerance)
< 150 mA
rms