Fuse Replacement
Main
ta
inin
g the SurgiS
ta
t
II
SurgiStat II-20 User’s Guide
5
-
3
Fuse Replacement
Fuses for the unit reside directly below the Power Cable Receptacle on the rear of
the unit.
To replace the fuses, follow this procedure:
1.
Unplug the power cord from the wall outlet.
2.
Remove the power cord from the Power Cable Receptacle on the rear panel.
3.
To release the fuse
drawer, insert a small
flathead screwdriver into
the slot on the drawer
below the power cord
receptacle and slide the
drawer out.
4.
Remove the two fuses and
replace them with new
fuses with the same values.
5.
Insert the fuse drawer into the Power Cable Receptacle.
Use the following fuses:
Surg II-20
110–120 V
VAC
250
Amps
5.0 A
Type
Slow Blow
Size
5 x 20 mm
Fuse
Drawer