24
Local
COMMAND
Returns meter to local mode.
Syntax:
Local<crlf>
Returns: <crlf>
Power-on default = LOCAL mode
Notes: REMOTE mode is selected when the meter receives a valid character
(not <crlf>.
LogOn
COMMAND
Turns on continuous reading mode synonymous with “Ohms?”. Note: if the
devices is already Logging, the LogOn command will return “Invalid
Command”
Syntax:
LogOn<crlf>
Returns:
Measurement Value + Multiplier every 300ms.
Example for 2
Ω
Range:
1.2345
Example for 20
Ω
Range:
12.345
Example for 200
Ω
Range:
123.45
Example for 2k
Ω
Range:
1.2345k
Example for 20k
Ω
Range:
12.345k
Example for 200k
Ω
Range:
123.45k
Example for 2M
Ω
Range:
1.2345M
Example for 20M
Ω
Range:
12.345M
LogOn0
COMMAND
Turns on continuous reading mode synonymous with “Ohms?”. Note: if the
devices is already Logging, the LogOn0 command will return “Invalid
Command”
Syntax:
LogOn0<crlf>
Returns:
Measurement Value + Multiplier every 500ms.
Example for 2
Ω
Range:
1.2345
Example for 20
Ω
Range:
12.345
Example for 200
Ω
Range:
123.45
Example for 2k
Ω
Range:
1.2345k
Example for 20k
Ω
Range:
12.345k
Example for 200k
Ω
Range:
123.45k
Example for 2M
Ω
Range:
1.2345M
Example for 20M
Ω
Range:
12.345M