Calibration Procedure
To perform calibration, you will need access to the main board of the device. Remove the top cover
screens from the front panel and all four screws from the rear panel. Remove the top cover and replace
the bottom two rear panel screws.
VOLTAGE SENSE ADJUSTMENTS
Connection:
Apply voltage at Sense Inputs and short Source and Sense negative Terminals.
On the main board press the CalSwitch (SW202) tact switch to initiate calibration routine.
4314Ki Procedure:
1.
Display will prompt:
2.
Press SW202. Display will prompt:
3.
Apply 0.000mV to the Voltage Sense and press SW202 on the Main Board.
Display will prompt:
4.
Apply -0.5mV and press SW202.
Display will prompt: