59
LOCAL COMMAND
Returns meter to local mode, remote LED off, Goto Local.
Syntax:
LOCAL
Returns:
<crlf>
Power-on default = LOCAL mode
Notes: REMOTE mode is selected when the meter receives a valid character (not <crlf>.
Once selected, all keys are disabled and will not be scanned. Therefore no key beeps with
the exclusion of the LOCAL key at the top right of the key pad. This key is active in REMOTE
mode and will extinguish the REMOTE LED and transition the device to the front panel
local user mode.
LOG COMMAND - Selects LOG mode on or off
Syntax :
LOG <ON or OFF>,0
(synonymous with OHMS?)
Example:
LOG ON 0<crlf>
(space or comma)
13.693<crlf>
13.664<crlf>
LOG ON,4
(default)
13.693<crlf>
13.664<crlf>
LOG ON,6
13.693,23.2,.0003931,13.699,20.0
Possible options, responses as if the following queries sent
===============
0 – OHMS?
1 – OHMS?,ATOD1?
2 – ATOD1?
3 – RDNG?
7 – EXTEMP?
8 – EXTEMP?,RDNG? (Run TCM ON first)
9 – LOG PENDING, starts when user presses PRINT
Power-on default = OFF