30
Setting the Limits
Each range has a different set of limits that can be programmed by the user. Table 6 lists the default values
programmed for each range. Once the user sets new values, the default settings will be overwritten, and
the user settings will be stored in memory.
To set your custom readings, simply select the range that better fits your load and follow this procedure.
Press
“Func. + HLC”
to bring-up the first input screen.
Using the numeric keys input a value for the upper limit. (Use
“C
LEAR
”
as a backspace to correct errors.)
Press
“E
NTER
”
to continue.
Using the numeric keys input a value for the lower limit. (Use
“C
LEAR
”
as a backspace to correct errors.)
Press
“E
NTER
”
to continue.
The screen will display a brief message and the instrument will
return to the standard measurement mode.
Table 6 –
HLC Default Limits
Range
Lower Limit
Upper Limit
2mΩ
1.0000mΩ
2.0000mΩ
20mΩ
10.000mΩ
20.000mΩ
200mΩ
100.00mΩ
200.00mΩ
2Ω
1.0000Ω
2.0000Ω
20Ω
10.000Ω
20.000Ω
200Ω
100.00Ω
200.00Ω
2kΩ
1.0000kΩ
2.0000kΩ
20kΩ
10.000kΩ
20.000kΩ