Section 4 | Functional Check
©
2023
–
Valeport Ltd
Page | 12
4
Functional Check
To perform a functional test connect the Hyperion to both power and PC using the supplied Y lead.
Run up the Configure App and check that data is being received in the Terminal window. The data
display will be in the selected output format, at the rate selected in the sampling option:
Now check the following:
The LEDs are on and light is being emitted from the fibres.
The magnitude of the data received increases when the supplied fluorescent target is held at 45° to
the sensor window:
By altering the angle of the target, the magnitude of the reading should change thereby showing
the correct operation of the sensor.
If the test target is missing then a piece of good quality white paper can be used.
4.1
User Calibration Factor and Offset