Hyperion Optical Sensors - Operating Manual:
MANUAL-360093108-14 | issue: 2.1
Page | 23
©
2023
–
Valeport Ltd
# Code followed by
Description
Type
Modbus
Register
Hex
Modbus
Register
Dec
Number of
Modbus
2 byte registers
Reply timing @
19200 Baud
(in ms)
#096
;gain;offset<CR><LF>
Set offset and gain coefficient for a user
secondary calibration optics channel 2 (when
used)
F_EXP
0x0037
55
4
<40
#097 <CR><LF>
Read offset and gain coefficient
F_EXP
0x0037
55
4
<40
#098
;modbus_allow_hash_
command<CR><LF>
Set to 1 if a # command is detected by the
Modbus system and setup mode entered
F_USHORT
0x0027
39
1
<40
#099
<CR><LF>
Read # recognition status
F_USHORT
0x0027
39
1
<40
#126
<CR><LF>
Read optical parameter id
F_UCHAR
0x0015
21
1
<40
#128
<CR><LF>
Read units
F_ASCII
0x0016
22
5
<40
#134
<CR><LF>
Read Serial number of the optical head
assembly
F_ULONG
0x0018
24
2
<40
#200
;operation_sequence<CR><LF>
Sets Operational Configuration
Normally automatically set by Macro commands
like M16 or C02 or S002
F_UCHAR
0x0009
9
4
<40
#201
<CR><LF>
Read setting for #200
F_UCHAR
0x0009
9
4
<40
#433
<CR><LF>
Read Background light intensity warning
level.
Light level that can cause loss of
accuracy/quality
F_UCHAR
0x0005
5
1
<40
#437
<CR><LF>
Read Time for LED stabilisation (inhibits output) F_UCHAR
0x0033
51
1
<40
#700
;modbus_enabled<CR><LF>
Set Modbus communication mode and parity,
Modbus mode
0;0 = disabled
1;0 = enabled
–
No Parity
1;1 = enabled
–
Even Parity
F_UCHAR
0x002C
44
1
<40
#701
modbus_enabled;<CR><LF>
Read Modbus communication mode and parity
F_UCHAR
0x002C
44
1
<40
#702
;uart_parity<CR><LF>
Set 0 for No Parity and 1 for Even parity in all
Uart communication (ascii and Modbus)
F_USHORT
0x002D
45
1
<40
#703
uart_parity;<CR><LF>
Read Parity for all UART communication (ascii
and Modbus)
F_USHORT
0x002D
45
1
<40