Hyperion Optical Sensors - Operating Manual:
MANUAL-360093108-14 | issue: 2.1
Page | 25
©
2023
–
Valeport Ltd
8
Electrical
Voltage:
9 - 28V DC isolated
Power:
40mA @ 12V DC
8.1
Interface Cable
–
0901EA2
END 1: 6 Way Male SubConn
SCMCIL6M.5
END 3: 7Way Male Souriau
UTS6JC10E7P
FUNCTION
PIN
PIN
1
G
RS485 GND
2
C
Tx RS232 out of instrument | RS485A(-)
3
D
Rx RS232 into Instrument | )
4
B
+V (9 - 28 V)
5
E
Select 232/485
Link to Pin G for RS485
6
A
-V
-
F
Screen
Note: to enable RS485 comms, link pins 1 and 5.
8.2
0400 Series Interface Cable
END 1:
Hyperion SUBCON
MCBH6F
END 2: 0400513
(Quartzonix Interface PCB)
END 3:
0400550 (PSU
Switching PCB)
FUNCTION
PIN
CONNECTOR
PIN CONNECTOR
1
J3, 6 WAY FCI
2
RS232 GND
2
3
RS232 RX (in to logger)
3
1
RS232 TX (out of logger)
4
J2, 2 way FCI
Hyperion Power +’ve
5
NO CONNECTION
(Blank off to prevent
shorting)
6
J2, 2 way FCI
Hyperion Power