![Vaisala Indigo 500 Series Скачать руководство пользователя страница 68](http://html2.mh-extra.com/html/vaisala/indigo-500-series/indigo-500-series_user-manual_3387257068.webp)
9.4 Restoring factory default settings
You can restore all device settings to the factory default settings. This will take some minutes.
A factory reset deletes all current settings of the device. After the
factory reset, you need to reconfigure the settings, including outputs and relays.
When you connect to the web interface the next time, you will be prompted to
give the activation code and create new users.
CAUTION!
Hex key (4 mm)
1. Power off the transmitter.
2. Open the 2 hex screws on the cover of the transmitter using a 4-mm hex key.
3. Open the transmitter cover.
4. Keep pressing the
PRESS TO ACTIVATE WEB INTERFACE
button while you power up the
transmitter. Keep pressing the button until the LED next to the button lights up.
ANALOG OUTPUTS
RELA
Y 1
RELA
Y 2
+
+
+
+
+
+
+
+
+
CH1
CH2
CH3
CH4
RS-485
PROBE
24 VOUT
ANALOG
INPUT
NO COM NC
NO COM NC
1
2
SERVICE PORT
PRESS TO ACTIVATE
WEB INTERFACE
ETHERNET
VS
+
The transmitter restarts.
If the transmitter asks you to calibrate the touchscreen display after the restart,
touch the requested crosshair positions on the touchscreen carefully and as
accurately as possible.
When the transmitter restarts after a factory reset, it will switch to using DHCP
and may get a new IP address. If needed, reconfigure the network settings using
the touchscreen. To continue using the web interface, reconnect using the
transmitter's new IP address.
Indigo 520 User Guide
M212287EN-A
66
Содержание Indigo 500 Series
Страница 1: ...M212287EN A User Guide Indigo 500 Series Transmitters Indigo 520...
Страница 5: ...Maintenance and calibration 77 Warranty 77 Technical support 77 Recycling 77 Table of contents 3...
Страница 8: ...Indigo 520 User Guide M212287EN A 6...
Страница 80: ......
Страница 81: ......
Страница 82: ...www vaisala com...