1
2
3
4
Figure 10 M20×1.5 cable gland with split
bushing
1
Base of the cable gland
2
Nylon seal
3
Split bushing inside the seal
4
Nut of the cable gland
3.3.6 Probe connection terminals and lead-throughs
Before connecting wires or cables, make sure that the transmitter is powered off.
ANALOG OUTPUTS
RELA
Y 1
RELA
Y 2
+
+
+
+
+
+
+
+
+
CH1
CH2
CH3
CH4
RS-485
PROBE
24 VOUT
ANALOG
INPUT
NO COM NC
NO COM NC
1
2
VS
+
1
2
For the M16×1.5 cable glands ordered
together with the transmitter from Vaisala,
the cable diameter is 2.0 … 6.0 mm
(0.08 … 0.24 in). Tightening torque for the
cable gland is 6 Nm.
The recommended maximum length of the
probe cable is 30 m (98 ft).
1
Probe 1 lead-through, M16×1.5
2
Probe 2 lead-through, M16×1.5
Indigo 520 User Guide
M212287EN-A
28