ERRS Command Response
Corresponding Error
Ambient temperature error
Ambient temperature out of range. [48]
Program flash check sum error
Firmware checksum mismatch. [49]
Parameter flash check sum error
Device settings corrupted. [50]
INFOA check sum error
Additional configuration settings corrupted. [51]
SCOEFS check sum error
Sensor coefficients corrupted. [52]
CURRENT check sum error
Main configuration settings corrupted. [53]
Voltage error
Supply voltage out of range. [55]
General flash failure w/r
Non-volatile memory read/write failure. [57]
Calibration certificate check sum
failure
Calibration certificate checksum mismatch. [58]
Refer to
Error Messages in Insight Software (page 70)
for more information about the error
states.
Table 9 Help Command
Syntax
Description
help
<cr>
Show list of currently available serial commands.
Chapter 6 – Using Service Port With Terminal Application