9.2 Error Messages in Insight Software
Table 15 Error Messages in Insight Software
Error
Likely Cause
Suggested Solution
Temperature
measurement error.
[44]
Temperature sensor is
short circuited,
damaged, or missing.
Check that the legs of the temperature sensor
are not short circuited. Contact a Vaisala Service
Center if the sensor is damaged.
Humidity
measurement error.
[45]
Humidity sensor is wet,
damaged, or missing.
Wait for the humidity sensor to dry, or remove
the filter and gently dry the sensor and the filter
with clean instrument air. Contact a Vaisala
Service Center if the sensor is damaged.
Supply voltage out of
range. [55]
Supply voltage is too
low.
Check and correct the power supply and wiring.
Firmware checksum
mismatch. [49]
Internal transmitter
failure.
Power-cycle the device, and if necessary, restore
the factory settings. If the error remains, contact
a Vaisala Service Center.
Device settings
corrupted. [50]
Additional
configuration settings
corrupted. [51]
Sensor coefficients
corrupted. [52]
Main configuration
settings corrupted.
[53]
Non-volatile memory
read/write failure. [57]
In case of constant error, contact a Vaisala Service Center.
More Information
‣
9.3 Error Codes in MI70 Handheld Meter
If several errors are active at the same time, the MI70 shows the sum of currently active error codes.
The label "
hex
" might not be displayed in MI70.
Chapter 9 – Troubleshooting
67
Содержание DRYCAP DMT143
Страница 1: ...M211435EN J User Guide Vaisala DRYCAP Dew Point Transmitter DMT143...
Страница 85: ...Appendix A Modbus Reference 83...
Страница 86: ...DMT143 User Guide M211435EN J 84...
Страница 88: ...DMT143 User Guide M211435EN J 86...
Страница 89: ......
Страница 90: ...www vaisala com...