When using the DSC74, seal the threads of the quick connector or thread adapter carefully
with PTFE thread seal tape. Tighten with an open-end wrench.
4
3
1
2
5
Figure 12 DSC74 Sampling Cell with Accessories
1
Thread adapter type G3/8" - G1/2"
2
Thread adapter type G3/8" - G1/4"
3
Quick connector
4
Leak screw
5
Sampling cell body (DMT242SC)
4.6.4 DSC74B Two-Pressure Sampling Cell
The DSC74B and DSC74C sampling cells are enhanced versions of the DSC74. The body is
designed in a way that it is possible to measure at both system pressure and atmospheric
pressure.
The DSC74B limits the flow rate with a fixed leak screw. The flow is optimized for pressures 3 ...
10 barg. The fixed leak screw eliminates the risk of opening the leak screw fully by accident
and this way emptying a gas vessel. The maximum flow can be increased, if needed, by
removing the leak screw and adjusting the flow manually with the valve. Harmful gases can be
recovered by connecting a collection system at the outlet (not available from Vaisala).
In the basic operation of the DSC74B, the gas flows to the sensor from the front and the outlet
is on the side. To have the measurement done at atmospheric pressure, the inlet and outlet are
reversed. Then the reducing parts supplied (G3/8" - G1/2" or G3/8" - G1/4") on the outlet side
help to protect the sensor from ambient humidity coming in.
DMT143 User Guide
M211435EN-J
30