Refer to
Error Messages in Insight Software (page 67)
for more information about the error states.
Table 16 Error Codes in MI70
Transmitter Error Code in MI70
Corresponding Error
1
hex
Temperature measurement error. [44]
2
hex
Humidity measurement error. [45]
40
hex
Device settings corrupted. [50]
80
hex
Additional configuration settings corrupted. [51]
100
hex
Sensor coefficients corrupted. [52]
200
hex
Main configuration settings corrupted. [53]
800
hex
Supply voltage out of range. [55]
10000
hex
Non-volatile memory read/write failure. [57]
20000
hex
Firmware checksum mismatch. [49]
9.4 Unknown Serial Settings
You can check the current serial settings of your transmitter by connecting to the serial line, or
with Vaisala Insight software.
More Information
‣
Accessing Serial Interface in Devices with Modbus Output (page 42)
‣
Vaisala Insight Software (page 36)
DMT143 User Guide
M211435EN-J