For FCC Compliance and product warnings/cautions, refer to your Software Administrator Guide.
Smart Sense Transmitters
The Smart Sense transmitter is worn on an infant’s ankle and
senses when the transmitter is loose or no longer in contact with the
infant.
The Smart Sense technology combines the Smart Sense
transmitter and Smart Sense banding material to give you the
capability to recognize if a transmitter has been removed from an
infant or if it has been tampered with or is loose.
Transmitters have a 1 year life expectancy. Using a transmitter
beyond the printed expiration date can result in system failure,
elopement, and/or abduction.
How do I band the transmitter?
1. Remove the Smart Sense banding material (with
the plastic guide tabs still attached) from the
plastic backing.
2. Open the clamps of the transmitter by gently
pulling them outward, one clamp at a time, from
the top of the transmitter.
3. With the yellow-side of the banding material
facing outwards, thread the banding into the first
clamp from the bottom slot out through the top
slot.
4. Once the plastic guide tab clears the clamp, snap
the first clamp closed.
5. Place the transmitter with banding around infant’s
ankle then thread the second clamp.
6. Snap the second clamp closed.
7. The transmitter becomes active 20 seconds after
the clamps are snapped into place to allow the
transmitter to be re-opened and adjusted.
8. Remove the plastic guide tabs and discard.
9. Use safety scissors to cut off the excess banding
material. Try not to cut bands too short to allow
room to adjust in the future.
WARNING
: Only Smart Sense banding material can be used to fasten
Smart Sense transmitters. Use of any other banding material with the
Smart Sense transmitters voids the RF Technologies' Product Warranty
and could result in system failure.
NOTE
: Do not thread both sides of the
banding material through the clamps until
you are ready to secure the transmitter to
the infant. Pre-threading may allow the
transmitter to alarm.