Disable heating before performing the verifier test.
1. Slip the verifier over the 3 transducers.
2. To see the wind speed measurement, establish a terminal connection, and give the
command
TABLE
.
3. Make sure that WMT700 reads less than 0.22 m/s (0.5 mph) with the verifier in place.
4. Remove the verifier.
5.7.2 Adding WMT700 to System
Before configuring the new sensor, complete the mechanical installation.
1. In AWS Client, select
Settings > Parameters > Static
.
2. In the
CFG
parameter set, select
WMT700_1
or
WMT700_2
.
3. To enable the sensor, type
1
in the
Value
field, and select
Save
.
4. Set the wind direction offset.
a. In the
General Settings
parameter set, select
wdoffset_1
or
wdoffset_2
.
b. In the
Value
field, type the wind direction offset in degrees.
c. Select
Save
.
AWS310-SITE Configuration and Maintenance Manual
M211828EN-C
102
Содержание AWS310-SITE
Страница 1: ...M211828EN C Configuration and Maintenance Manual Vaisala Automatic Weather Station AWS310 SITE ...
Страница 10: ...AWS310 SITE Configuration and Maintenance Manual M211828EN C 8 ...
Страница 21: ...Chapter 2 Product Overview 19 ...
Страница 22: ...AWS310 SITE Configuration and Maintenance Manual M211828EN C 20 ...
Страница 62: ...AWS310 SITE Configuration and Maintenance Manual M211828EN C 60 ...
Страница 86: ...AWS310 SITE Configuration and Maintenance Manual M211828EN C 84 ...
Страница 140: ...AWS310 SITE Configuration and Maintenance Manual M211828EN C 138 ...
Страница 148: ...AWS310 SITE Configuration and Maintenance Manual M211828EN C 146 ...
Страница 150: ...AWS310 SITE Configuration and Maintenance Manual M211828EN C 148 ...
Страница 152: ...AWS310 SITE Configuration and Maintenance Manual M211828EN C 150 ...
Страница 166: ...AWS310 SITE Configuration and Maintenance Manual M211828EN C 164 ...
Страница 172: ...AWS310 SITE Configuration and Maintenance Manual M211828EN C 170 ...
Страница 174: ...AWS310 SITE Configuration and Maintenance Manual M211828EN C 172 ...
Страница 175: ......
Страница 176: ...www vaisala com ...