Installation and connection
Set up the vacuum pump
5 cm
5 cm
5 cm
>
Place the vacuum pump on a stable, non‑vibrating, level
surface.
>
When installing in lab furniture, maintain a minimum dis
‑
tance of 5 cm (2 in) to adjacent objects or surfaces.
>
The product must be positioned so that the on/off switch
and the power plug can be reached and are accessible. For
this purpose a minimum distance of 12 cm (5 in) to adjacent
objects or surfaces must be kept at one side of the pump.
The on/off switch is at the rear of the terminal box.
>
Prevent heat accumulation and ensure sufficient air
circulation, especially in closed housings.
Observe limitation of use
Limitation of use
(US)
Operating ambient temperature
12 – 40 °C
54 – 104 °F
Max. altitude
2000 m
above sea level
6500 ft
above sea level
Minimum distance to
adjacent parts
5 cm (12 cm)
2 in (5 in)
Relative humidity
30
–
85 %, non‑condensing
Pollution degree
2
Protection class
IP 40
type 1
Prevent condensation or external contamination from dust, liquids, and
corrosive gases.
Example
Sketch of
minimum distances
in lab furniture
IMPORTANT!
Содержание RE 2.5
Страница 1: ...Instructions for use OI no 20901559 Original instructions EN RE 2 5 RZ 2 5 RE 6 RZ 6 Rotary Vane Pump ...
Страница 24: ...24 20901559_EN_RE_RZ_V7_230523 Introduction ...
Страница 40: ...40 20901559_EN_RE_RZ_V7_230523 Installation and connection ...
Страница 48: ...48 20901559_EN_RE_RZ_V7_230523 Installation and connection ...
Страница 73: ...73 20901559_EN_RE_RZ_V7_230523 Appendix 8 8 8 8 CU Certificate CU Certificate CU Certificate ...
Страница 76: ...76 20901559_EN_RE_RZ_V7_230523 ...
Страница 77: ...77 20901559_EN_RE_RZ_V7_230523 ...
Страница 78: ...78 20901559_EN_RE_RZ_V7_230523 ...
Страница 79: ...79 20901559_EN_RE_RZ_V7_230523 ...