Important notes regarding the use of gas ballast
Gas ballast is a continuous purge to keep the pump’s in-
terior as clean as possible and to reduce the possibility of
condensation inside the pump.
➨
Air and pumped media might react inside the pump or
at the outlet of the pump and form hazardous or explo-
sive mixtures, when you use air rather than inert gas
for the gas ballast. This constitutes a risk of significant
damage to equipment and/or facilities, a risk of per-
sonal injury or even loss of life.
+
Make sure that air/gas intake through the gas ballast
valve can never lead to hazardous, explosive or other-
wise dangerous mixtures. If in doubt, use inert gas.
To reduce condensation in the pump, do not pump vapor
before the pump has reached its operating temperature.
Open the gas ballast valve when pumping condensable
vapors. Turn gas ballast cap to open valve.
NOTICE
gas ballast
(open)
For
condensable vapors
(water vapor, sol-
vents, etc.):
- The gas ballast valve is open if the arrow on
the gas ballast cap is pointing towards the
inlet of the pump.
- With gas ballast valve open, the ultimate
vacuum will be reduced.
- Use inert gas for gas ballast to avoid the formation of
explosive mixtures. A special adapter fitting is needed
to connect an inert gas supply line (see “Accessories
/ Spare parts”, pg. 43). This adapter replaces the
standard gas ballast cap and allows for an inert gas
line to be connected via a KF DN 16 small flange at a
maximum supply pressure of 17.5 psi (1.2 bar) abso-
lute.
- Close the gas ballast valve by turning the cap 180°.
Содержание ME 16C NT-ME 16C NT+EK
Страница 29: ...page 29 of 88 ME 16C NT MD 12C NT MV 10C NT 1 5 3 2 6 4 8 7 9...
Страница 49: ...page 49 of 88 Disassembling the pump housing 1 3 2...
Страница 50: ...page 50 of 88 4 6 5 1 2 4 4x 2 5 mm 2 5 mm...
Страница 51: ...page 51 of 88 7 9 8 4 4x without EK...
Страница 52: ...page 52 of 88 10 12 11 4 2x 4 4x Don t loosen this screw...
Страница 53: ...page 53 of 88 13 15 14 5 2x 5 2x...
Страница 54: ...page 54 of 88 16 18 17 2 5 mm...
Страница 56: ...page 56 of 88 Replacing the diaphragms 19 21 20 Clean 5 8x 20696867 20696839...
Страница 57: ...page 57 of 88 22 24 23 1 2 3 Pay attention to number of washers...
Страница 58: ...page 58 of 88 25 27 26...
Страница 61: ...page 61 of 88 34 36 35 Clean 20696867 20696839...
Страница 62: ...page 62 of 88 37 39 38 IN IN EX EX 4x...
Страница 63: ...page 63 of 88 40 42 41 IN EX T20 4x 2 2 ft lbf 3 Nm...
Страница 64: ...page 64 of 88 43 45A 44 Maintain all pump heads in the same way MD 12C NT...
Страница 65: ...page 65 of 88 45B 45C 45D ME 16C NT MV 10C NT ME 16C NT MD 12C NT MV 10C NT...
Страница 67: ...page 67 of 88 4A 3 2 size 2 2x Version A valve 24 4 mm...
Страница 68: ...page 68 of 88 5A 20696839 24 4 mm 26 mm 4B 5B 20696839 37 mm 40 mm Version B valve 37 mm...
Страница 69: ...page 69 of 88 6 7 size 2 2x 7A without EK size 2 2x...
Страница 71: ...page 71 of 88 Assembling the pump housing 46 47 48 5 2x 8 9 ft lbf 12 Nm...
Страница 72: ...page 72 of 88 50 49 5 2x 8 9 ft lbf 12 Nm 51...
Страница 73: ...page 73 of 88 53 52 4 4x 3 7 ft lbf 5 Nm 54 4 2x 3 7 ft lbf 5 Nm...
Страница 74: ...page 74 of 88 56 55 4 4x 3 7 ft lbf 5 Nm 57...
Страница 75: ...page 75 of 88 59 58 60 4 4x 3 7 ft lbf 5 Nm...
Страница 76: ...page 76 of 88 62 61 63...
Страница 79: ...page 79 of 88 4 20638821 20638821 5 6...
Страница 80: ...page 80 of 88 7 T20 4x size 1 4x 8...
Страница 83: ...page 83 of 88 Declarations and certificates...
Страница 84: ...page 84 of 88...
Страница 87: ...page 87 of 88...