(chemicals, thermal decomposition products of fluoro
-
elastomers). Use appropriate protective clothing and
safety goggles.
+
Interruption of the pump (e.g., due to power failure),
failure of connected components or of parts of the sup-
ply, or change in parameters must not be allowed to
lead to dangerous conditions. In case of a diaphragm
failure or in case of a leak in the manifold, pumped
substances might be released into the environment or
into the pump housing or motor.
Comply with all notes regarding proper use of the pumps,
as well as operation and maintenance guidance.
+
The residual
leak rate of the equipment
might render
possible an exchange of gas, albeit extremely slight,
between the environment and the vacuum system.
Adopt suitable measures to prevent contamination of
the pumped substances or the environment.
• Ensure that no parts of your clothing, hair or fingers
can be caught or drawn in at the inlet of the pump.
Never insert fingers or drop any other object into the
inlet or outlet.
• Pumping at high inlet pressure may lead to overpres-
sure at the gas ballast valve. Pumped gases or con-
densate might be expelled if the valve is open. If an in-
ert gas supply is connected to the gas ballast, ensure
that its inlet pipeline is not contaminated.
•
Pay attention to the safety symbol ”hot surfaces” on
the equipment. Depending on operation conditions and
ambient conditions dangers due to hot surfaces may
arise. Hot parts may cause burns if touched. Adopt
suitable measures to prevent any danger arising from
hot surfaces. Ensure that hot surfaces of the pump do
not cause burns. Provide a suitable contact guard if
necessary.
The surface temperature of the glass components at
Содержание ME 16C NT-ME 16C NT+EK
Страница 29: ...page 29 of 88 ME 16C NT MD 12C NT MV 10C NT 1 5 3 2 6 4 8 7 9...
Страница 49: ...page 49 of 88 Disassembling the pump housing 1 3 2...
Страница 50: ...page 50 of 88 4 6 5 1 2 4 4x 2 5 mm 2 5 mm...
Страница 51: ...page 51 of 88 7 9 8 4 4x without EK...
Страница 52: ...page 52 of 88 10 12 11 4 2x 4 4x Don t loosen this screw...
Страница 53: ...page 53 of 88 13 15 14 5 2x 5 2x...
Страница 54: ...page 54 of 88 16 18 17 2 5 mm...
Страница 56: ...page 56 of 88 Replacing the diaphragms 19 21 20 Clean 5 8x 20696867 20696839...
Страница 57: ...page 57 of 88 22 24 23 1 2 3 Pay attention to number of washers...
Страница 58: ...page 58 of 88 25 27 26...
Страница 61: ...page 61 of 88 34 36 35 Clean 20696867 20696839...
Страница 62: ...page 62 of 88 37 39 38 IN IN EX EX 4x...
Страница 63: ...page 63 of 88 40 42 41 IN EX T20 4x 2 2 ft lbf 3 Nm...
Страница 64: ...page 64 of 88 43 45A 44 Maintain all pump heads in the same way MD 12C NT...
Страница 65: ...page 65 of 88 45B 45C 45D ME 16C NT MV 10C NT ME 16C NT MD 12C NT MV 10C NT...
Страница 67: ...page 67 of 88 4A 3 2 size 2 2x Version A valve 24 4 mm...
Страница 68: ...page 68 of 88 5A 20696839 24 4 mm 26 mm 4B 5B 20696839 37 mm 40 mm Version B valve 37 mm...
Страница 69: ...page 69 of 88 6 7 size 2 2x 7A without EK size 2 2x...
Страница 71: ...page 71 of 88 Assembling the pump housing 46 47 48 5 2x 8 9 ft lbf 12 Nm...
Страница 72: ...page 72 of 88 50 49 5 2x 8 9 ft lbf 12 Nm 51...
Страница 73: ...page 73 of 88 53 52 4 4x 3 7 ft lbf 5 Nm 54 4 2x 3 7 ft lbf 5 Nm...
Страница 74: ...page 74 of 88 56 55 4 4x 3 7 ft lbf 5 Nm 57...
Страница 75: ...page 75 of 88 59 58 60 4 4x 3 7 ft lbf 5 Nm...
Страница 76: ...page 76 of 88 62 61 63...
Страница 79: ...page 79 of 88 4 20638821 20638821 5 6...
Страница 80: ...page 80 of 88 7 T20 4x size 1 4x 8...
Страница 83: ...page 83 of 88 Declarations and certificates...
Страница 84: ...page 84 of 88...
Страница 87: ...page 87 of 88...