page 30 of 70
outlet tubing must be at least the size of the
pump’s exhaust connection.
• Reduce the transmission of vibration. Prevent mechan-
ical load due to rigid pipelines. Insert elastic hoses or
flexible elements as couplings between the pump and
rigid pipes.
Secure hose connections at the pump ap-
propriately, e.g., with hose clamps, to protect
against accidental detachment.
EX
✔
✘
Always install outlet tubing descending from
the pump to avoid backflow of condensate to-
wards the pump.
Plug in the power cord.
• Check the power source and the pump’s
rating plate to be sure that the power source
and the equipment match in voltage, phase,
and frequency.
Switch the pump on.
O
➨ I
NOTICE
NOTICE
Electrical connection
Содержание ME 16 NT
Страница 42: ...page 42 of 70 3 4 5 4 4x 4 4x 4 2x...
Страница 43: ...page 43 of 70 Don t loosen this screw 7 8 6 5 2x...
Страница 44: ...page 44 of 70 9 10 11 5 2x...
Страница 46: ...page 46 of 70 12 2 5 mm 13 14 2 5 mm 5 8x...
Страница 47: ...page 47 of 70 15 696819 696827 16 17 Clean...
Страница 48: ...page 48 of 70 696819 696827 18 19...
Страница 50: ...page 50 of 70 Assembling the pump head 23 24 25 Pay attention to number of washers...
Страница 52: ...page 52 of 70 29 30 31...
Страница 53: ...page 53 of 70 32 5 8x Screw in diagonally at first slightly then tighten 8 9 ft lbf 12 Nm 33 34...
Страница 54: ...page 54 of 70 35 Maintain all pump heads in the same way ME 16 NT MD 12 NT MV 10 NT 36A 36B...
Страница 55: ...page 55 of 70 Replacing the valve at the distributor only MD 12 NT MV 10 NT 1 2 size 2 2x 2 5 mm 3 size 2 2x...
Страница 56: ...page 56 of 70 4 5 6 696827...
Страница 57: ...page 57 of 70 7 8 size 2 2x 9 size 2 2x...
Страница 59: ...page 59 of 70 Assembling the pump housing 37 38 39 5 2x 8 9 ft lbf 12 Nm...
Страница 60: ...page 60 of 70 40 42 41 5 2x 8 9 ft lbf 12 Nm...
Страница 61: ...page 61 of 70 45 4 2x 3 7 ft lbf 5 Nm 44 4 4x 3 7 ft lbf 5 Nm 43...