page 17 of 70
of such substances from the pump outlet, install an
appropriate collecting and disposal system and take
protective action for pump and environment.
➨
You must take suitable precautions to prevent any for-
mation of explosive mixtures in the pump chamber or
at the outlet of the pump. In case, e.g., of a diaphragm
failure, mechanically generated sparks, hot surfaces
or static electricity may ignite these mixtures. Use inert
gas for venting, if necessary.
➨
Drain appropriately or otherwise remove any poten-
tially explosive mixtures at the outlet of the pump, or
dilute them to non-explosive concentrations.
➨
Never operate this pump if it has a damaged cord or
plug.
+
If the pump is not working properly, has been dropped
or has fallen into water, contact your pump service pro-
vider.
+
Prevent any part of the human body from coming into
contact with vacuum.
+
Always provide a free and pressureless exhaust outlet
to avoid damage to pump valves and risk of bursting.
+
Attention
: Due to clogging of the silencer an internal
overpressure may build up, which may damage bear-
ings, diaphragms and valves. Preferably use the hose
nozzle as outlet and make sure that the exhaust pipe-
line cannot become blocked.
+
Comply with applicable regulations when disposing of
chemicals. Take into consideration that chemicals may
be contaminated. Take adequate precautions to pro-
tect people from the effects of dangerous substances
(chemicals, thermal decomposition products of fluoro-
elastomers). Use appropriate protective clothing and
safety goggles.
Содержание ME 16 NT
Страница 42: ...page 42 of 70 3 4 5 4 4x 4 4x 4 2x...
Страница 43: ...page 43 of 70 Don t loosen this screw 7 8 6 5 2x...
Страница 44: ...page 44 of 70 9 10 11 5 2x...
Страница 46: ...page 46 of 70 12 2 5 mm 13 14 2 5 mm 5 8x...
Страница 47: ...page 47 of 70 15 696819 696827 16 17 Clean...
Страница 48: ...page 48 of 70 696819 696827 18 19...
Страница 50: ...page 50 of 70 Assembling the pump head 23 24 25 Pay attention to number of washers...
Страница 52: ...page 52 of 70 29 30 31...
Страница 53: ...page 53 of 70 32 5 8x Screw in diagonally at first slightly then tighten 8 9 ft lbf 12 Nm 33 34...
Страница 54: ...page 54 of 70 35 Maintain all pump heads in the same way ME 16 NT MD 12 NT MV 10 NT 36A 36B...
Страница 55: ...page 55 of 70 Replacing the valve at the distributor only MD 12 NT MV 10 NT 1 2 size 2 2x 2 5 mm 3 size 2 2x...
Страница 56: ...page 56 of 70 4 5 6 696827...
Страница 57: ...page 57 of 70 7 8 size 2 2x 9 size 2 2x...
Страница 59: ...page 59 of 70 Assembling the pump housing 37 38 39 5 2x 8 9 ft lbf 12 Nm...
Страница 60: ...page 60 of 70 40 42 41 5 2x 8 9 ft lbf 12 Nm...
Страница 61: ...page 61 of 70 45 4 2x 3 7 ft lbf 5 Nm 44 4 4x 3 7 ft lbf 5 Nm 43...
Страница 62: ...page 62 of 70 46 4 4x 3 7 ft lbf 5 Nm 47 4 4x 3 7 ft lbf 5 Nm 48...