Take adequate precautions to protect people from the
effects of dangerous substances if contamination has
occurred. Use appropriate protective clothing, safety
goggles and protective gloves.
+
Wear parts have to be replaced regularly.
+
Never operate a defective or damaged pump.
+
Check every motor capacitor regularly by measuring
its capacity and estimating its time in operation. Re-
place old capacitors early enough to prevent a failure
in operation. If an old motor capacitor fails, the capaci-
tor may get hot. It may even melt or emit a flame, which
could be
dangerous for persons and equipment in
the vicinity
. The capacitors have to be replaced by an
electrician.
+
Vent the pump before starting maintenance. Isolate the
pump and other components from the vacuum system.
Allow sufficient cooling of the pump. Drain condensate,
if applicable.
Ensure that
maintenance
is done only by suitably trained
and supervised technicians. Ensure that the maintenance
technician is familiar with the safety procedures, which re-
late to the products processed by the pumping system.
Only dismantle the pump as far as necessary.
NOTICE
Содержание ME 16 NT
Страница 42: ...page 42 of 70 3 4 5 4 4x 4 4x 4 2x...
Страница 43: ...page 43 of 70 Don t loosen this screw 7 8 6 5 2x...
Страница 44: ...page 44 of 70 9 10 11 5 2x...
Страница 46: ...page 46 of 70 12 2 5 mm 13 14 2 5 mm 5 8x...
Страница 47: ...page 47 of 70 15 696819 696827 16 17 Clean...
Страница 48: ...page 48 of 70 696819 696827 18 19...
Страница 50: ...page 50 of 70 Assembling the pump head 23 24 25 Pay attention to number of washers...
Страница 52: ...page 52 of 70 29 30 31...
Страница 53: ...page 53 of 70 32 5 8x Screw in diagonally at first slightly then tighten 8 9 ft lbf 12 Nm 33 34...
Страница 54: ...page 54 of 70 35 Maintain all pump heads in the same way ME 16 NT MD 12 NT MV 10 NT 36A 36B...
Страница 55: ...page 55 of 70 Replacing the valve at the distributor only MD 12 NT MV 10 NT 1 2 size 2 2x 2 5 mm 3 size 2 2x...
Страница 56: ...page 56 of 70 4 5 6 696827...
Страница 57: ...page 57 of 70 7 8 size 2 2x 9 size 2 2x...
Страница 59: ...page 59 of 70 Assembling the pump housing 37 38 39 5 2x 8 9 ft lbf 12 Nm...
Страница 60: ...page 60 of 70 40 42 41 5 2x 8 9 ft lbf 12 Nm...
Страница 61: ...page 61 of 70 45 4 2x 3 7 ft lbf 5 Nm 44 4 4x 3 7 ft lbf 5 Nm 43...
Страница 62: ...page 62 of 70 46 4 4x 3 7 ft lbf 5 Nm 47 4 4x 3 7 ft lbf 5 Nm 48...