38
20901121_12+16_VARIO_Vsel_V1.0_020818
Commissioning (operation)
Process screen
1
2
3
4
5
6
1
Status bar
2
Analogue pressure display – pressure curve
3
Digital pressure reading – pressure value
(target value, actual value, pressure unit)
4
Process screen with context features
5
Screen navigation
6
Operating elements for control
Operating elements
Button
Function
active
locked
Start
`
Start application – only available on the process
screen.
Stop
`
Stop application – always possible.
*
VENT
– vent the system (option)
`
Press button < 2 sec = vent briefly; control con‑
tinues.
*
`
Press button > 2 sec = vent to atmospheric pres‑
sure; vacuum pump is stopped.
`
Press button during venting = venting is stopped.
Favorites
`
View Favorites menu.
* Button is only displayed if venting valve is connected or activated.
Process screen
vacuum controller
Vacuum controller
operating elements