27
20901121_12+16_VARIO_Vsel_V1.0_020818
Installation and connection
Installing the vacuum pump
>
Place the vacuum pump on a stable, non‑vibrating, level sur‑
face.
5 cm
5 cm
5 cm
>
When installing in lab furniture, maintain a minimum distance
of 5 cm (2 in.) to adjacent objects or surfaces.
>
Prevent heat accumulation and ensure sufficient air circula‑
tion, especially in closed housings.
Observe limitations of use
Limitation of use
(US)
Ambient temperature
10
–
40 °C
50–104°F
Max. altitude
2000 m
above NHN
6562 ft
above sea level
Minimum distance to
adjacent parts
5 cm
2 in
Relative humidity
30–85 %, non condensing
Protection class
IP 40/IK 08
Prevent condensation or contamination from dust, liquids, or corrosive
gases.
>
Note the IP protection class. IP protection is only guaranteed
if the device is appropriately mounted and connected.
>
For connection also note the rating plate data and chapter
8.1.1 Technical data on page 70.
IMPORTANT!
Example
Sketch of
minimum distances
in laboratory
furniture
Observe limitation of
use
IMPORTANT!