Commissioning (operation)
5.3 Shutdown (switch off)
Take the pump out of operation
1.
Stop the process.
2.
Disconnect the pump from the apparatus.
3.
Let the vacuum pump run on for about 30 minutes, with open
inlet.
5
Condensate and media residues will be flushed out of the
vacuum pump.
IMPORTANT!
>
Prevent deposits and rinse condensate out of
the pump.
2
(a)
4.
Switch rocker switch
(a)
off – switch position
0
.
5
Pump switched off.
5.
Check the pump for dirt and damage.
Switch pump off