Safety information!
Ensure that repair is done only by a suitable trained and supervised technician. Obey
local and national safety requirements. Ensure that the technician is familiar with the
safety procedures which relate to the substances processed by the pumping sys-
tem. Ensure that the device is decontaminated before repair and that you take ad-
equate precautions to protect people from the effects of dangerous substances if
contamination has occurred.
Read the instructions for use for the equipment carefully, especially sections “Safety
informations“ and “Use and operation“. If necessary order ”Instructions for use” from
VACUUBRAND.
If hazardous or corrosive substances have been used:
☞
The device will be contaminated with the process chemicals which have been
pumped during operation. Ensure that the device is decontaminated before main-
tenance.
☞
Take adequate precautions to protect people from the effects of dangerous sub-
stances. Wear appropriate safety-clothing, do not inhale and avoid contact with
skin.
Before starting repair, isolate the device and other components from the vacuum
system and the
electrical supply
so that they
cannot be operated accidentally.
Allow the device to cool, so that it is at safe temperature for maintenance work.
Before starting maintenance,
wait two minutes
after isolating the equipment from
mains to allow the capacitors to discharge.
Before starting repair check that the required parts are available and of the correct
type before you start your work.
☞
Check the operating sequence mentally on feasibility, safety requirements and
consequences on safety and function of the equipment.
Use only genuine spare parts and accessories.
☞
If using components of other manufacturers the function and the safety of the
equipment may be restricted.
☞
Do not reuse damaged parts.
Check operability and safety after repair.
☞
In case of problems return the device to the factory for inspection or repair if
necessary.
Obey regulations when disposing of solvents and chemicals.
In order to comply with law (occupational, health and safety regulations, safety at
work law and regulations for environmental protection) vacuum pumps, components
and measuring instruments returned to the manufacturer can be repaired only when
certain procedures (see section “Notes on return to the factory“ in the ”Instructions
for use”) are followed.
Scrapping and waste disposal
Dispose of the pump and any components removed from it safely in accordance with
all local and national safety and environmental requirements.
Particular care must be taken with components and waste oil which have been con-
taminated with dangerous substances from the process.
☞
You may authorize us to dispose the equipment at your expense.