background image

Seite 7 von 12 // page 7 of 12

Das Dokument darf nur vollständig und unverändert verwendet und weitergegeben werden. Es liegt in der Verantwortung des
Anwenders, die Gültigkeit dieses Dokumentes bezüglich seines Produktes sicher zu stellen.  999195  //  05/02/2008

Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefully
the validity of this document with respect to his product.   999195  //  05/02/2008

Safety information!

Ensure that repair is done only by a suitable trained and supervised technician. Obey
local and national safety requirements. Ensure that the technician is familiar with the
safety procedures which relate to the substances processed by the pumping sys-
tem. Ensure that the device is decontaminated before repair and that you take ad-
equate precautions to protect people from the effects of dangerous substances if
contamination has occurred.

Read the instructions for use for the equipment carefully, especially sections “Safety
informations“ and “Use and operation“. If necessary order ”Instructions for use” from
VACUUBRAND.

If hazardous or corrosive substances have been used:

The device will be contaminated with the process chemicals which have been
pumped during operation. Ensure that the device is decontaminated before main-
tenance.

Take adequate precautions to protect people from the effects of dangerous sub-
stances. Wear appropriate safety-clothing, do not inhale and avoid contact with
skin.

Before starting repair, isolate the device and other components from the vacuum
system and the 

electrical supply

 so that they 

cannot be operated accidentally.

Allow the device to cool, so that it is at safe temperature for maintenance work.
Before starting maintenance, 

wait two minutes

 after isolating the equipment from

mains to allow the capacitors to discharge.

Before starting repair check that the required parts are available and of the correct
type before you start your work.

Check the operating sequence mentally on feasibility, safety requirements and
consequences on safety and function of the equipment.

Use only genuine spare parts and accessories.

If using components of other manufacturers the function and the safety of the
equipment may be restricted.

Do not reuse damaged parts.

Check operability and safety after repair.

In case of problems return the device to the factory for inspection or repair if
necessary.

Obey regulations when disposing of solvents and chemicals.

In order to comply with law (occupational, health and safety regulations, safety at
work law and regulations for environmental protection) vacuum pumps, components
and measuring instruments returned to the manufacturer can be repaired only when
certain procedures (see section “Notes on return to the factory“ in the ”Instructions
for use”) are followed.

Scrapping and waste disposal
Dispose of the pump and any components removed from it safely in accordance with
all local and national safety and environmental requirements.
Particular care must be taken with components and waste oil which have been con-
taminated with dangerous substances from the process.

You may authorize us to dispose the equipment at your expense.

NOTICE

NOTICE

NOTICE

Содержание CVC 3000

Страница 1: ...ch seines Produktes sicher zu stellen 999195 05 02 2008 Documents are only to be used and distributed completely and unchanged It is strictly the users responsibility to check carefully the validity o...

Страница 2: ...t abk hlen lassen Sicherstellen dass das Ger t keinesfalls im ge ffneten Zustand unbeabsichtigt anlaufen kann Vor jedem Eingriff nach Trennen der Ger te vom Netz zwei Minuten warten bis sich die Kond...

Страница 3: ...y the users responsibility to check carefully the validity of this document with respect to his product 999195 05 02 2008 Vor Beginn der Arbeiten Abschnitt Unbedingt beachten lesen Netzstecker ziehen...

Страница 4: ...carefully the validity of this document with respect to his product 999195 05 02 2008 Schlauchclip am Schlauch des Einlassblocks mit einem Schraubendreher mit Flachklinge ffnen Schraubendreher wie in...

Страница 5: ...lauchenden stecken Schlauch zum Controller auf den unteren Schlauchanschluss Schlauch zur Pumpe auf den oberen Schlauchanschluss des neuen Einlassblocks stecken Schlauchclips mit einer Zange schlie en...

Страница 6: ...essertes Enddruckverhalten bei Fl ssigkeitsanfall Der Rundkolben ist au en beschichtet Splitterschutz bei Implosion Auslaufschutz bei mechani scher Besch digung Bei Kondensatanfall Den Fl ssigkeitssta...

Страница 7: ...r isolate the device and other components from the vacuum system and the electrical supply so that they cannot be operated accidentally Allow the device to cool so that it is at safe temperature for m...

Страница 8: ...users responsibility to check carefully the validity of this document with respect to his product 999195 05 02 2008 Before starting read and obey section Safety information Isolate pumping unit from...

Страница 9: ...he users responsibility to check carefully the validity of this document with respect to his product 999195 05 02 2008 Remove both screws at the counterholder at the inlet block using Torx screwdriver...

Страница 10: ...8 Pull controller PC 3001 basic out of the stand housing Unplug connection cable Assemble the catchpot at the inlet using joint clip Position hose clips at the ends of the hoses Connect the hose to th...

Страница 11: ...s of diaphragms and valves are enhanced Improves vacuum in case of condensation The round bottom flasks are coated with a protective layer to prevent disintegration in case of break age or implosion I...

Страница 12: ...h unserem Einfluss entziehen Wir bitten deshalb um Verst ndnis dass aus unserer Beratung keine Anspr che abgeleitet werden k nnen Die bertragbarkeit ist daher im Einzelfall vom Anwender selbst sehr so...

Отзывы: