1. INSTALLATION
Because of metrological legislation, installation/some
metrological parameter settings are limited to be done
by authorized personnel only. Do not attempt to change
any of the built-in parameters. Contact your dealer for
installation and technical assistance.
CAUTION:
This unit is legal for trade only when it is sealed (and/or
stamped) and bearing a serial number. Do not attempt
to break the seal (or stamp) affixed to this indicator
or remove the serial number. Contact your dealer for
more information and after sales service.
For most accurate weighing result, do not use the
indicator in where or when the environment condition
falls beyond as those listed on
SPECIFICATIONS
.
Do not attempt to open this unit or conduct any trouble
shootings other than those listed on
TROUBLE SHOOTING
.
2. SPECIFICATIONS
Model No.
ABM/AFM/AFS/ATM INDICATOR
Max.
USER DEFINED
n
max
3000
(NOTE 1)
Tare Range
1/3 Max. ¡ e
(NOTE 2)
Power-on Zero Range
¡Ó10% Max.
Zero Range
¡Ó2% Max.
Min. Load
20e
(NOTE 3)
Max. measuring range
9.0mV
Min. signal voltage per
verification scale interval
1.5 V
Power voltage requirements
6.0 VDC (BY EXTERNAL POWER
ADAPTOR OR RECHARGEABLE
BATTERY)
Minimum battery voltage
5.5 VDC
Load cell Excitation voltage
5 VDC
Minimal load cell impedance
85
Maximal load cell impedance
1000
Load cell connection
4 wires
Operation
Environment
0
o
~40
o
C
Non-condensed. R.H.
¡
85%
Specifications subject to change without notice.
NOTE 1:
The recommend division when application is legal for
trade (OIML mode) is 3000 and 6000 to 7500 when
application is not legal for trade (normal mode). The
maximum display division under OIML mode when Extended
Display function engaged is 30000, the same as under
non-OIML mode.