![UWE ABM Скачать руководство пользователя страница 10](http://html1.mh-extra.com/html/uwe/abm/abm_operation-manual_1066077010.webp)
transaction to memory
c. Indicator displays ¡
P. X
¡¨ and
M+
INDICATOR
appears
to indicate that memory contains of stored data
NOTE
: ¡
P. X
¡¨ means the total number of transactions
accumulated to memory.
d. Indicator returns to normal display status after 2
seconds
e. Repeat
a
to
c
for subsequent transactions
NOTE:
- When scale is set at OIML and application is legal
for trade. User only needs to press
M+
once when trying
to store a value to memory. The weight accumulation
function will take place when reading is stabled
- When scale is set at Normal and application is not
legal for trade. User can save the shown value to memory
at anytime when pressing
M+
5.8.2 Memory Recall
Press
MR
to recall total accumulated weight from memory.
After
MR
is pressed, indicator displays
P. X
(
X
Means
the number of transactions accumulated) follow by a value.
The value is the total accumulated weight stored in
memory.
When the
total accumulated weight is being displayed,
MR INDICATOR
appears to indicator that the value being
displayed is the total accumulated weight. Indicator
will return to normal display status after 3 seconds.
5.8.3 Memory Clear
All transactions stored in memory by pressing
MR
and
TARE
at the same time.
CAUTION:
All data stored will be erased when the indicator
is turned/powered off.
5.9 EXTENDED DISPLAY
This indicator is equipped with extended display for
reference when it is operating under OIML mode. Press
and hold
MODE
for 2 seconds and display will show the
extended digit for 5 seconds before return to normal.
5.10 RECHARGE SCALE
The indicator is equipped with a built-in rechargeable
battery. When the
LO-BAT INDICATOR
appears, recharge
the scale immediately. Failure to do so may cause
unrecoverable damage to the rechargeable battery.
Charging is completed when the color of the
IN-CHARGE
INDICATOR
turns green.