![Ushio UIT-250 Скачать руководство пользователя страница 24](http://html1.mh-extra.com/html/ushio/uit-250/uit-250_operating-manual_1064224024.webp)
18
1000033244-00
4. Instructions for use
- Reference table 3 below concerning irradiance values and temperature when
the analog voltage signal outputs 1 V.
Table 3
Irradiance value and temperature when analog signal output is 1 V
Model
UVD-C254
UVD-S254
(mW/cm
2
)
UVD-C365
UVD-S365
(mW/cm
2
)
UVD-C405
UVD-S405
(mW/cm
2
)
UVD-TK
(
°
C)
H range
9999
9999
999.9
M range
999.9
999.9
99.99
Indication
limits
L range
99.99
99.99
9.999
200.0
4-3-2. Output of irradiance (temperature) distribution
Irradiance and temperature distribution is stored inside the UIT-250 (maximum
2 minutes (32 samples/second) or 4 minutes (16 samples/second) for 3840 pieces of
data) and is output using the analog voltage signal.
If placed on the conveyor of the UV radiation hardening equipment with a conveyor
(UNICURE) the irradiance distribution or temperature distribution can be measured.
(1) Perform the "4-2-5. Accumulated exposure amount measurement (1) to (5).
- There measurement range can not be changed if a temperature sensor is con-
nected.
(2) Connect the analog signal output cable that came with the unit to the analog
signal output terminal.
- Connect the orange and white wires of the analog signal output cable respec-
tively to the +,
−
terminals of the recorder device (oscilloscope, pen recorder
etc.).
Important
1. The output impedance of the analog signal output is 1 k
Ω
. In order to
provide a correct measurement it is recommended that the input im-
pedance of the recorder device be greater than 1 M
Ω
.
2. If the input impedance of the recorder device is low, the measured
value may be low.
Содержание UIT-250
Страница 1: ...USHIOINC System Company...
Страница 2: ......
Страница 39: ......
Страница 40: ...USHIOINC URL http www ushio co jp...