![Ushio UIT-250 Скачать руководство пользователя страница 14](http://html1.mh-extra.com/html/ushio/uit-250/uit-250_operating-manual_1064224014.webp)
8
1000033244-00
4. Instructions for use
(4) Using the precision driver that came with the unit, fix the photodetector to the
UIT-250 upper back side using the 2 screws that came with the unit.
(5) When using a separate-type photodetector, connect a separate photodetector
where the model and serial number match that of the sensitivity adjustment
module.
(6) When using a temperature sensor, confirm that the model of the sensitivity
adapter is "UVD-TK" and connect it to the temperature sensor section.
•
Removal of the photodetector and/or sensitivity adjustment module
(1) Remove 2 screws from the upper back side of the UIT-250 using the precision
driver that came with the unit.
(2) Cause the photodetector to face the front of the exposure meter and rotate the
photodetector.
(3) When the fold portion of the photodetector is removed from the body, the pho-
todetector can be removed.