FPMAN-1-190214
Page 5
GENERAL SPECIFICATIONS
Flow Ranges:
2 SLPM/5 SCFH F.S. to 1300 SLPM/2600 SCFH F.S.
Turndown Ratio:
400:1
Accuracy:
± 1% of reading
Repeatability:
± 0.2% of full-scale
Maximum Measurable Flow:
Up to 125% of full-scale, gas dependent
Pressure Effect on Accuracy:
Less than
±
0.04% of reading/ PSI
Temperature Effect on Accuracy:
Less than
±
0.04% of reading/ °F
Response Time:
100 msec
Pressure Drop:
2.5 PSI at maximum flow (port to port), gas dependent
Gases:
Air, Argon, Nitrogen, CO2, Oxygen, Helium, Hydrogen,
Methane, and user selectable
Gas compatibility:
Non-corrosive, non-condensing
Maximum Operating Pressure:
150 PSIG
Sensor Burst Pressure:
200 PSIG
Operating Temperature:
-25 to 80 °C (-13 to 176 °F)
Process Connections:
1/8”-1/4”-3/8”-1/2”-3/4” NPT. SAE, BSPT, BSPP also available
Display:
Rate, total, pressure, temperature, multi-gas, alarms, engineering
units
Wetted Parts:
Sensors: Ceramic, silicon, gold, epoxy, RTV
Flow body: Stainless steel, anodized aluminum, Viton®
Enclosure Rating:
Type 4
ELECTRICAL SPECIFICATIONS
Analog output:
4-20 mA (2-wire loop powered)
Alarms:
2 independent open-collector outputs (high/low flow rate)
Open-Collector Rating:
30VDC at 50 mA
Electrical Connection:
5-pin connector
Supply Voltage:
10–30 VDC
Supply Current:
22 mA @ F.S. flow (includes over-range)