FPMAN-1-190214
Page 9
KEYPAD OPTIONS
UP:
Consecutive (short) button press cycles between the following information on top of the
display.
a.
Gas
: Current selected Gas
b.
Pressure
: Gas pressure in PSI
c.
Pressure
: Gas pressure in PSIA
d.
Totalizer
: Total volume of gas according to the selected units.
e.
Temperature
: Temperature in degree Celsius
f.
Temperature
: Temperature in degree Fahrenheit
Options are saved in the memory so that next time you turn the power on it will remember
the display settings.
DOWN:
Consecutive button press cycles between the different flow units in SCFH and SLPM.
Once selected, the setting is saved into the memory of the device.
To reset the totalizer, press
SET
until the LCD shows “TARING”. Release the SET button
immediately, and the totalizer will be reset.
To tare the meter (at zero flow) press
SET
until the LCD shows “TARING”. Keep holding the
SET button (for 5 seconds) until the LCD shows “TARED” and the flow output will be set to
zero.
5 seconds