7 TECHNICAL DATA
LCD Graphic Display
The LCD display is based on the Electronic Assembly EA DOGM132W-5 display module:
•
Resolution: 132x32 Pixels
•
Viewing Area: 48mm x 13mm
•
Text Color: Black
•
Background Color: White
•
Backlight: Configurable
Analog Signals
Thermocouple Input
Thermocouple Scale: NIST 175, ITS-90
Type K 0-1200°C, reference 0°C: 0 – 48828.0 uV
Type S 0-1760°C, reference 0°C: 0 – 18612.38 uV
Type R 0-1760°C, reference 0°C: 0 – 21006.36 uV
Type B 0-1820°C, reference 0°C: 0 – 13813.50 uV
•
Range: 0-100 mV
•
Resolution: 20 bit
•
Accuracy: 0.02% of FS (at 25’C)
•
Full Scale Temperature Coefficient: 10 ppm/°C
•
Input Impedance: >100 MΩ
•
Cold Junction Compensation: 0.03°C per °C
Thermocouple
Type
ºF
ºC
Zero
Span
Zero
Span
B
800
3000
420
1680
K
32
2300
0
1200
R
300
3000
150
1650
S
300
3000
150
1650
Thermocouple Output:
•
Range: 0-100 mV
•
Resolution: 16 bit
•
Accuracy: 0.05% of FS (at 25’C)
•
Full Scale Temperature Coefficient: 10 ppm/°C
•
Max Operating Current: 20mA
Millivolt Input:
•
Range: 0-2000mV
•
Resolution: 20 bit
•
Accuracy: 0.02% of FS (at 25’C)