Action
?
Check continuity of the DXBUS cable.
?
Check for shorts between the DXBUS pins. An
open or short circuit on another DX Modules can
cause this fault.
?
If the hazard lights were already switched on
when the DX system was turned on, Flash code
10 and limp Mode may result. To clear this fault,
turn the Hazard lights off, then turn the DX
System off then on again.
?
If generated by a Kill signal, the cause of the fault
is severe.
11
Stall Timeout Fault
Cause
The motor current has been at, or
close to, current limit for longer than
the Stall Timeout parameter value.
Action
?
Turn the DX System off then on
again.
12
Module Mismatch
Cause
There is a compatibility problem
between DX Modules in the system.
The wheelchair will be disabled.
Action
?
Consult your Dynamic service
center.
Cause
The data held by the DX UCM for
another DX Module is corrupt or
incompatible with that module
Action
?
Reprogramming the wheelchair
system may correct this problem.
Limp Mode
If the DX System detects some faults, it will revert to Limp Mode. This is a
reduced speed mode which recognises problems, but allows the wheelchair
user to limp home, where the problem can be assessed.