16
4
Torque Settings
16
Torque Settings
Chapter
4
4-1.
Calibration
"Calibration" is the operation of matching the TM400 with a sensor.
Check the UTM
Ⅱ
to be used.
4-2.
Calibration procedure for torque
Register the rated capacity value of the sensor.
Set the sign of the sensor.
Set the zero point of the sensor in an unloaded condition
(where no force and so on is applied to a sensor).
Set the minimum unit of torque such as scale interval
and scale division. (optional if there is no change)
UTM capacity (UTM CAP.)
Select the rated capacity of UTM
Ⅱ
connected.
<Setting range>
0.03, 0.05, 0.1, 0.2, 0.3, 0.5, 1, 2, 3, 5, 10, 20, 30, 50,100, 200,
300, 500, 1000, 2000, 3000, 5000, 10000 [Nm], OTHER
* The following setting items only available when OTHER is selected, and arbitrary calibration
can be made.
- Other capacity - Unit - Decimal point
Other capacity (OTHER CAP.)
(These items only available in case the UTM capacity is set to OTHER)
Set the arbitrary torque value against 5V input from UTM
Ⅱ
.
<Setting range>
1 to 99999
Unit (UNIT)
(These items only available in case the UTM capacity is set to OTHER)
Set the unit of torque to display.
<Setting range>
mNm, Nm, kNm
4
Torque Settings
(1)
UTM capacity
*
*
*
:
Be sure to perform the setting.
(3)
Zero calibration
*
(4)
Minimum scale
(2)
Sign